Internal Quality Control Program (IQC) for Neonatal Phe Screening Test
Phe IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC1502AC | TPC1502AD |
ReagentKit | PerkinElmer-GSP |
Unit | mg/dL blood |
Duration of the Analyzing | Month (2023/12) | CUM (2018/04/01~2023/12/31) | Month (2023/12) | CUM (2018/04/01~2023/12/31) |
Runs (N) | 9 | 927 | 9 | 927 |
Mean | 2.0 | 2.3 | 7.5 | 9.7 |
SD | 0.20 | 0.73 | 0.86 | 3.26 |
CV (%) | 10.0 | 31.7 | 11.5 | 33.6 |
Target Value | 1.8 | 1.8 | 7.5 | 7.5 |
Total Error (%) | 31.1 | 91.3 | 22.9 | 96.5 |
TEa (%) | 13 | 13 | 13 | 13 |
σ | 0.2 | -0.5 | 1.1 | -0.5 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2018-04-01 to 2023-12-31 ; Target : 1.8 ; SD : 0.23 ( ●) | |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2018-04-01 to 2023-12-31 ; Target : 7.5 ; SD : 0.98 ( □) | |
|
Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2018-04-01 to 2023-12-31 ; Target : 1.8 ; SD : 0.23 | |
|
Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2018-04-01 to 2023-12-31 ; Target : 7.5 ; SD : 0.98 | |
|
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 29 Mar, 2022