Internal Quality Control Program (IQC) for Neonatal Phe Screening Test
Phe IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC1502AC | TPC1502AD |
ReagentKit | PerkinElmer-GSP |
Unit | mg/dL blood |
Duration of the Analyzing | Month (2024/06) | CUM (2018/04/01~2024/06/30) | Month (2024/06) | CUM (2018/04/01~2024/06/30) |
Runs (N) | 8 | 935 | 8 | 935 |
Mean | 1.9 | 2.3 | 8.1 | 9.7 |
SD | 0.22 | 0.73 | 0.79 | 3.25 |
CV (%) | 11.6 | 31.7 | 9.8 | 33.5 |
Target Value | 2.0 | 2.0 | 7.4 | 7.4 |
Total Error (%) | 28.2 | 78.5 | 29.0 | 98.1 |
TEa (%) | 13 | 13 | 13 | 13 |
σ | 0.7 | -0.1 | 0.4 | -0.5 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 2.0 ; SD : 0.36 ( ●) | |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 7.4 ; SD : 0.76 ( □) | |
 |
Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 2.0 ; SD : 0.36 | |
 |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 7.4 ; SD : 0.76 | |
 |