Phe 缺乏症篩檢檢驗之實驗室內部品管計畫
Internal Quality Control Program (IQC) for Neonatal Phe Screening Test
線上回報系統 (Phe IQC NS MIS System) [登入/login]
財團法人預防醫學基金會 執行
| CI | CII |
QC Control Lot No. | TPC2101AG | TPC2101AH |
ReagentKit | Guangzhou Fenghua Bionegineering CO. LTD. |
Unit | mg/dL blood |
檢驗期間 | Month (2023/12) | CUM (2024/07/01~2024/09/30) | Month (2023/12) | CUM (2024/07/01~2024/09/30) |
累計點數 (N) | 0 | 66 | 0 | 66 |
累計平均值 (Mean) | - | 3.0 | - | 10.1 |
累計標準差 (SD) | - | 0.21 | - | 0.76 |
累計變異係數 (CV) % | - | 7.0 | - | 7.5 |
目標值 (Target) | 2.9 | 2.9 | 10.3 | 10.3 |
總誤差 (TE) % | - | 17.4 | - | 17.0 |
可容許總誤差 (TEa) % | 30 | 30 | 30 | 30 |
σ | - | 3.8 | - | 3.7 |
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: mg/dL blood
檢驗期間: 2024-07-01 至 2024-09-30 ; 目標值 (Target) : 2.9 ; 參考標準差 (SD) : 0.29 ( ●) | |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: mg/dL blood
檢驗期間: 2024-07-01 至 2024-09-30 ; 目標值 (Target) : 10.3 ; 參考標準差 (SD) : 1.03 ( □) | |
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : mg/dL blood
檢驗期間: 2024-07-01 至 2024-09-30 ; 目標值 (Target) : 2.9 ; 參考標準差 (SD) : 0.29 | |
|
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : mg/dL blood
檢驗期間: 2024-07-01 至 2024-09-30 ; 目標值 (Target) : 10.3 ; 參考標準差 (SD) : 1.03 | |
|