Phe 缺乏症篩檢檢驗之實驗室內部品管計畫
Internal Quality Control Program (IQC) for Neonatal Phe Screening Test
線上回報系統 (Phe IQC NS MIS System) [登入/login]
財團法人預防醫學基金會 執行
| CI | CII |
QC Control Lot No. | TPC2101AG | TPC2101AH |
ReagentKit | Guangzhou Fenghua Bionegineering CO. LTD. |
Unit | mg/dL blood |
檢驗期間 | Month (2025/03) | CUM (2024/07/01~2025/03/31) | Month (2025/03) | CUM (2024/07/01~2025/03/31) |
累計點數 (N) | 5 | 120 | 5 | 120 |
累計平均值 (Mean) | 2.8 | 3.0 | 10.4 | 10.2 |
累計標準差 (SD) | 0.32 | 0.22 | 0.63 | 0.73 |
累計變異係數 (CV) % | 11.4 | 7.3 | 6.1 | 7.2 |
目標值 (Target) | 2.9 | 2.9 | 10.3 | 10.3 |
總誤差 (TE) % | 26.3 | 18.1 | 13.1 | 15.3 |
可容許總誤差 (TEa) % | 30 | 30 | 30 | 30 |
σ | 2.3 | 3.6 | 4.8 | 4.0 |
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: mg/dL blood
檢驗期間: 2024-07-01 至 2025-03-31 ; 目標值 (Target) : 2.9 ; 參考標準差 (SD) : 0.29 ( ●) | |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: mg/dL blood
檢驗期間: 2024-07-01 至 2025-03-31 ; 目標值 (Target) : 10.3 ; 參考標準差 (SD) : 1.03 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : mg/dL blood
檢驗期間: 2024-07-01 至 2025-03-31 ; 目標值 (Target) : 2.9 ; 參考標準差 (SD) : 0.29 | |
 |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : mg/dL blood
檢驗期間: 2024-07-01 至 2025-03-31 ; 目標值 (Target) : 10.3 ; 參考標準差 (SD) : 1.03 | |
 |