Internal Quality Control Program (IQC) for Neonatal Phe Screening Test
Phe IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC1901AE | TPC1901AF |
ReagentKit | PerkinElmer-GSP |
Unit | mg/dL blood |
Duration of the Analyzing | Month (2023/08) | CUM (2020/06/01~2023/08/31) | Month (2023/08) | CUM (2020/06/01~2023/08/31) |
Runs (N) | 14 | 716 | 14 | 716 |
Mean | 1.7 | 2.0 | 6.4 | 7.2 |
SD | 0.17 | 0.29 | 0.47 | 0.77 |
CV (%) | 10.0 | 14.5 | 7.3 | 10.7 |
Target Value | 2.0 | 2.0 | 7.6 | 7.6 |
Total Error (%) | 35.0 | 29.0 | 30.5 | 26.7 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 0.5 | 1.4 | 0.6 | 1.4 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1901AE ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 2.0 ; SD : 0.35 ( ●) | |
Lot No.: TPC1901AF ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 7.6 ; SD : 0.51 ( □) | |
 |
Lot No.: TPC1901AE ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 2.0 ; SD : 0.35 | |
 |
Lot No.: TPC1901AF ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 7.6 ; SD : 0.51 | |
 |