Phe 缺乏症篩檢檢驗之實驗室內部品管計畫
Internal Quality Control Program (IQC) for Neonatal Phe Screening Test
線上回報系統 (Phe IQC NS MIS System) [登入/login]
財團法人預防醫學基金會 執行
| CI | CII |
QC Control Lot No. | TPC2101AG | TPC2101AH |
ReagentKit | Guangzhou Fenghua Bionegineering CO. LTD. |
Unit | mg/dL blood |
檢驗期間 | Month (2024/11) | CUM (2024/07/01~2024/11/30) | Month (2024/11) | CUM (2024/07/01~2024/11/30) |
累計點數 (N) | 5 | 96 | 5 | 96 |
累計平均值 (Mean) | 3.1 | 3.0 | 10.3 | 10.2 |
累計標準差 (SD) | 0.19 | 0.20 | 0.96 | 0.73 |
累計變異係數 (CV) % | 6.1 | 6.7 | 9.3 | 7.2 |
目標值 (Target) | 2.9 | 2.9 | 10.3 | 10.3 |
總誤差 (TE) % | 19.2 | 16.8 | 18.6 | 15.3 |
可容許總誤差 (TEa) % | 30 | 30 | 30 | 30 |
σ | 3.8 | 4.0 | 3.2 | 4.0 |
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: mg/dL blood
檢驗期間: 2024-07-01 至 2024-11-30 ; 目標值 (Target) : 2.9 ; 參考標準差 (SD) : 0.29 ( ●) | |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: mg/dL blood
檢驗期間: 2024-07-01 至 2024-11-30 ; 目標值 (Target) : 10.3 ; 參考標準差 (SD) : 1.03 ( □) | |
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : mg/dL blood
檢驗期間: 2024-07-01 至 2024-11-30 ; 目標值 (Target) : 2.9 ; 參考標準差 (SD) : 0.29 | |
|
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : mg/dL blood
檢驗期間: 2024-07-01 至 2024-11-30 ; 目標值 (Target) : 10.3 ; 參考標準差 (SD) : 1.03 | |
|
若有任何疑問,歡迎來電(函)聯繫或指教
地 址:10699 台北郵政 26-624 號信箱 電 話:(02)2703-6080
e-mail:
傳 真:
(02)2703-6070
財團法人預防醫學基金會
updated : 29 Mar, 2022