Internal Quality Control Program (IQC) for Neonatal Phe Screening Test
Phe IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC1502AC | TPC1502AD |
ReagentKit | PerkinElmer-GSP |
Unit | mg/dL blood |
Duration of the Analyzing | Month (2020/06) | CUM (2018/07/01~2020/06/30) | Month (2020/06) | CUM (2018/07/01~2020/06/30) |
Runs (N) | 1 | 434 | 1 | 434 |
Mean | 3.2 | 3.0 | 9.6 | 12.6 |
SD | - | 0.39 | - | 1.21 |
CV (%) | - | 13.0 | - | 9.6 |
Target Value | 2.9 | 2.9 | 12.3 | 12.3 |
Total Error (%) | - | 29.4 | - | 21.6 |
TEa (%) | 30 | 30 | 30 | 30 |
σ | - | 2.0 | - | 2.9 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 2.9 ; SD : 0.39 ( ●) | |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit: mg/dL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 12.3 ; SD : 1.07 ( □) | |
 |
Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 2.9 ; SD : 0.39 | |
 |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit : mg/dL blood
Duration : 2018-07-01 to 2020-06-30 ; Target : 12.3 ; SD : 1.07 | |
 |